Skip to main content

Stratyx250 System Specifications

Details about the 250mL controller hardware, operating ranges, utility requirements, and installation details.

Updated over 5 months ago

Installation Requirements

Space Requirements, Bay Cart

Specification

Detail

Notes

Min. access width

36”

Smallest possible door width between loading dock and final location

System dimensions

44” x 21” x 54”

Dimensions (WxDxH)

Required front clearance

24”

Space left vacant in front of the system for swinging open cabinet doors and tipping reactors forward during maintenance

Overall footprint depth

46”

Cart depth + maintenance clearance

Overall footprint width

44”

(excludes utility cart)

Cart weight

150.35

lbs

Single bay weight

45.15

lbs

Full assembly weight

330.95

lbs

Space Requirements, Utility Cart

Specification

Detail

Notes

System Dimensions

23” x 21” x 29”

Dimensions (WxDxH)

Gas inlets

4x push-to-connect

¼” tube OD, semi-firm tubing, rated 120 psi or greater

Power inlet

NEMA 5-20 plug

Sole Power connection to reactor system

Network inlet

CAT-6 wired ethernet

Chiller inlet

½” OD x 0.32” ID

Tubing must be fitted by installer, no quick-disconnects. Chiller must be within 4ft of the utility cart.

Weight

99.85

lbs

Space Requirements, Chiller

Specification

Detail

Notes

System Dimensions

14.2” x 23.6” x 24.4”

Dimensions (WxDxH)

Refrigerant

R134A

Reservoir capacity

2.5

Gallons

Cooling capacity

1000W

3415 BTU

Temperature range

-10°C to + 30°C

Stability

+/- 0.1°C

Weight

9

lbs

Space Requirements, Overall System

Specification

Detail

Notes

System footprint

4-bay

83” x 46” x 54”

System with (4) bioreactors, (1) utility cart, and (1) chiller

(W x D x H) includes clearance for maintenance

System footprint

8-bay

127” x 46” x 54”

System with (8) bioreactors, (1) utility cart, and (1) chiller

(W x D x H) includes clearance for maintenance

System footprint

12-bay

169” x 46” x 54”

System with (12) bioreactors, (1) utility cart, and (1) chiller

(W x D x H) includes clearance for maintenance

Overall weight

430.8 / 761.8 / 1180

4 / 8 / 12 bay system (lbs)

Requirements, Autofiller

Specification

Detail

Notes

System footprint

11” x 6.5” x 11.5”

Dimensions (WxDxH)

Network connection

WIFI 6E 802.11ax

Wired ethernet available as backup

Power

Standard NEMA 5-15, 120V receptacle

Max power usage 120 W

Weight

17.85

lbs

System Electrical Requirements

Specification

Detail

Notes

Circuit capacity

20A x 2

Dedicated 120V 20A circuit/breaker for the reactor system, separate circuit for standalone chiller*

Continuous power

500W / 1000W / 1500W

Max continuous power draw, for 4 / 8 / 12 reactor configurations

Line frequency

60 Hz

Line voltage

120 VAC

Plug format

NEMA 5-20

Wall outlet located within 10ft of utility cart

Environmental requirements

Specification

Detail

Notes

Lab temperature

15-35°C recommended

5-40°C max operating range

Humidity

< 75% RH, non-condensing

80% max, non-condensing

Elevation

<2000 m

Gas requirements

Specification

Detail

Notes

Air inlet pressure

30-100 psig

300 sccm/reactor

2.4 SLPM per system max flow

Nitrogen inlet pressure

30-100 psig

300 sccm/reactor

2.4 SLPM per system max flow

O2 inlet pressure

30-100 psig

100 sccm/reactor

1.2 SLPM per system max flow

CO2 inlet pressure

30-100 psig

100 sccm/reactor

1.2 SLPM per system max flow

Connections at utility cart

1/4" push-to-connect

Recommend ¼” OD semi-flexible gas tubing with inline cutoff valve

Gas sources

Connection to tank

Within 10’ of utility cart

Networking requirements

Specification

Detail

Notes

Bandwidth

100 Kbps per reactor

During experiment operation

Main system connection

Wired ethernet

WiFi is possible but not recommended at this time

Autofiller connection

WiFi

(optional: ethernet)

Physical network connection

RJ-45, CAT5/5e/6

Network switches

Managed (recommended)

Unmanaged switches are not recommended at this time

Inbound ports

None

No inbound connections

Outbound ports

80, 443

Operating and Control Specifications

Temperature control specifications

Specification

Detail

Temperature range

4 - 45 °C

Sensor accuracy

0.1 °C

Control accuracy

± 0.1 °C at 37 °C

Heating rate*

~ 0.5 °C / min

Cooling rate*

~ 0.5 °C / min

Temperature uniformity

0.1 °C

*Temperature dependent

Gas control specifications

Specification

Detail

Gas types

Air, N2, O2, CO2

MFC range

0 - 100 sccm (each)

Primary sparge mix

O2 + CO2 + (Air or N2)

Primary sparge position

Below first impeller

Primary sparge type

Open pipe, L-shape

Overlay

Air or N2

DO sensor type

Optical patch

DO range

0-100% up to 45mg/L O2

Agitation specifications

Specification

Detail

Impeller type

3-blade, elephant ear 45 deg angle

Impeller diameter

24mm

Impeller count

2

Impeller spacing

30mm

Baffle intrusion

6.5mm (~1/10 of vessel diameter)

Tip speed

0.5 m/s @400rpm

Agitation speed

0 - 1000 rpm

pH control specifications

Specification

Detail

pH sensor type

Optical patch

(optional single-use glass)

pH range (patch)*

5.5 - 8.5

pH range (glass, optional)*

2 - 12

Control strategies

Acid, base, and CO2. One-sided or two-sided with adjustable deadbands.

Control accuracy**

~ 0.05 - 0.25 pH

*see vessel spec sheet for more detailed sensor specifications

**control accuracy depends on control strategy and PID parameters

Fluid delivery specifications

Specification

Detail

Pump type

Syringe / positive displacement

Pump qty

5

Syringe capacity (mL)

100, 50, 10, 50, 100

Flow rate range

5 µL/min - 50 mL/min

Flow rate accuracy

± 5% at > 0.1 mL/min

± 0.005 mL/min at < 0.1 mL/min

Bolus volume accuracy

± 5% at > 0.40 mL

± 0.020 mL at < 0.40 mL

Expansion port

The expansion port on the front face of the bay is a generic M12 connection designed to support a broad array of compatible PAT (Process Analytical Technology) solutions. The Culture vessel supports a standard 120mm long probe with PG13.5 threading. Please discuss with your product specialist to find out more about supported probes and new probe development options.

Did this answer your question?